العنوان
Perianaesthetic blood glucose measurement ;
الناشر
Mona Saied Ahmed Ahmady,
المؤلف
Ahmady, Mona Saied Ahmed.
هيئة الاعداد
باحث / Mona Saied Ahmed Ahmady
مشرف / Mohamed Mohamed Atallah
مشرف / Golnar Seddik Hammouda
مشرف / Sameh Mokhitar Banob
الموضوع
Blood Glucose-- metabolism.
تاريخ النشر
1995.
عدد الصفحات
29 p. ;
اللغة
الإنجليزية
الدرجة
ماجستير
التخصص
التخدير و علاج الألم
تاريخ الإجازة
1/1/1995
مكان الإجازة
جامعة المنصورة - كلية الطب - DEPARTMENT OF ANESTHESIA
الفهرس
Only 14 pages are availabe for public view

from 41

from 41

Abstract

This study has been planned to evaluate the accuracy of glucometer (Accutrend). Fourty blood glucose measurements included in this study, they were randomly selected. Patients were anaesthetized by different techniques. Blood glucose concentration was measured by both glucometer and the standard laboratory glucose-oxidase method (Beckman Autoanalyse r).
from the results of this study, blood glucose measured by glucometer (Accutrend) are always lower than those measured by the standard lab glucose oxidase method and the calculated bias value was 17.6 ± 51.8 and the limits of agreement were wide ranging from 121.2 to - 86 mg dl-1 which are
unaccepted for clinical purposes.